Burst Volume and Burst Pressure Test

Application
For measuring burst volume and burst pressure simultaneously according to ISO 4074: 2002 Standard - Annex G

Description of the machine

  • 2-6 independent cabinets, each cabinet has one mount for one condom     
  • Air Flow accuracy ± 1%      
  • Volume accuracy ± 1%      
  • Pressure accuracy of pressure transmitter max. ± 0.1% pressure range 0-5 kPa.       
  • Sound insulation at the walls, noise absorbent easy to clean plastic surface
  • Powder coated solid steel sheet and frame construction
  • Automatic door closing during test and opening after test the other sides are noise insulated
     
  • Volume metering system based on reliable and accurate gas meter
  • Newly designed automatic test head for easy unrolling of condoms and auto load of clamping mechanism – cup places itself automatically
  • Break-test and abort-test button shown and calculated also in the test reports
     
  • Recording of the measured results can be stored on PC and printed out in test forms, when desired
  • Air flow and -pressure are adjustable at three installed pressure regulator valves
  • All modular technique – for easy exchange of spar parts and reduced  maintenance cost – the equipment needs no special maintenance except cleaning
  • Including precision calibration equipment one Rota Meter and one Manometer
  • Pentium 4 PC, 17” monitor, licensed Windows based laboratory software
  • Spare parts for 2 years of operation
  • Number of cabinets to connect to PC is up to eight units
     
  • Electronic and Software made in Germany
  • Free service and upgrade are on line available

Technical data:
Power supply: 220/240V, 50/60Hz
Method: Compressed air 6 bar, oil and water free
Weight: c. 75 kg per cabinet
Dimensions: 550mm x 550mm x 1,400mm per cabinet